EX-1 Single Channel & EX4-400 Quad Channel
Differential Extracellular Amplifiers (CORNERSTONE Series)
![]()
|
The EX4-400 and EX1 Amplifiers are ideal for all non-human applications which require single or multiple channel AC or DC amplification with High and Low frequency filtering. With the optional differential headstage, the EX4-400 and EX1 are perfect single or multi-channel extracellular amplifier systems with gains of 10 to 20,000. |
Features
Specifications
|
Mainframe Amplifier |
Headstage |
||
Standard: + 700mV Optional: + 2V Optional: + 10V
Low Cut: .1, .3, 1, 2, 3, 5, 10, 20, 30, 100, 300 High Cut: 30, 100, 300, 500, 1k, 2k, 3k, 5k, 10k, 30k, 100k
EX4-400: Includes PS-1 External Power Supply 115/230VAC 50/60Hz EX1: Includes PS-2 Externally Mount Power Supply 115VAC 60Hz only(For 230VAC, order PS-1 Upgrade)
EX1: 7.3" wide, 2.6" high, 9.7" deep.
|
|
||
|
Source Resistance = Shorted Input |
|||
|
Bandwidth |
RMS |
P-P |
|
|
1Hz -100Hz |
0.3uV |
1.5uV |
|
|
1Hz -1KHz |
0.7uV |
3.5uV |
|
|
1Hz -10KHz |
1.6uV |
8.0uV |
|
|
Source Resistance = 100 Kilohm |
|||
|
Bandwidth |
RMS |
P-P |
|
|
1Hz -100Hz |
0.55uV |
2.75uV |
|
|
1Hz -1KHz |
1.6uV |
8.0uV |
|
|
1Hz -10KHz |
3.9uV |
19.5uV |
|
|
Source Resistance = 1 Megohm |
|||
|
Bandwidth |
RMS |
P-P |
|
|
1Hz -100Hz |
1.6uV |
8.0uV |
|
|
1Hz -1KHz |
4.7uV |
23.5uV |
|
|
1Hz -10KHz |
11uV |
55uV |
|
|
Source Resistance = 10 Megohm |
|||
|
Bandwidth |
RMS |
P-P |
|
|
1Hz -100Hz |
5.0uV |
25uV |
|
|
1Hz -1KHz |
14uV |
70uV |
|
|
1Hz -10KHz |
25uV |
125uV |
|
|
PROBE INPUT CONNECTORS (please
specify one of the two below) The 4001 or 4002
Headstages come with
BNC connector or 1mm (0.04") Tip
Jacks Mates with Dagan HB120 and HB180 microelectrode holders for glass pipettes. Can also be used with BNC pins with metal electrodes.<> <>1 mm Tip Jacks (0.04" dia) Mates with standard Tip Jack connectors By special order a min coaxial cable as used on older 2400's is available with a $70 surcharge. |
Options:
4001 Differential Headstage: Gain of 10, with Capacitance Neutralization (normally used with glass electrodes)
4002 Differential Headstage: Gain of 10 (normally used with metal electrodes)
Integrator (True RMS): Back panel BNC provides true RMS output of amplified signal.
Sample & Hold: TTL activated Sample & Hold via back panel BNC to reduce artifacts.
Auto Zero Offset: Tracking Integrator control located on back panel See Description
Line Notch Filter: Filters line frequency noise from output signal (specify 50 or 60 Hz).
Relay Probe: BNC and switch in small box (on probe cable), with break-away relay, for use in staining.
|
EX1 |
One channel of the EX4-400 in a bench top cabinet (includes wall mount power supply: 115V AC only) |
$ 920.00 |
|
EX4-400 |
Quad Differential Amplifier (Includes External Power Supply for 115 or 230VAC) |
$ 3,160.00 |
|
4001 |
Extracellular Differential Headstage, Gain of 10, with cap. comp. |
$ 450.00 |
|
4002 |
Extracellular Differential Headstage, Gain of 10, without cap. comp. |
$ 350.00 |
|
PS-1 |
Line operated Power Supply required for 230 V for EX1 |
Add $ 90.00 |
|
EX-I |
Integrator (True RMS) per channel -- |
Add $95.00 |
|
EX-SO |
Sample and Hold per channel -- |
Add $75.00 |
|
EX-AO |
Auto Zero Offset Option See Description |
Add $300.00 |
|
EX-Notch |
50/60Hz Line Notch Filter Option -per channel -- |
Add $125.00 |
|
RP |
Current Injection Relay Probe (add $200.00 to the Headstage price) |
Add $200.00 |
| See the Holders and Accesories section for mating connectors, jacks and cables |
Warning: This equipment is not intended for use in human applications or human experimentation.
[ Home | News | Products | Ordering | Feedback ]
Dagan Corporation Minneapolis MN USA 612-827-5959 (FAX 612-827-6535) Email: info@dagan.com 1/08