EX-1000 (Cornerstone Series)

10 Channel Extracellular Amplifier

 

 

Press HERE to view full size front panel (file size 30K)

 

The EX-1000 Extracellular Amplifier is ideal for all non-human applications that require multiple channel AC amplification. Each channel is totally independent with variable Gains, High Cut and Low Cut Filters. Inputs for all 10 channels are through a single high impedance headstage.

Features

 

Specifications

  • Input Connector:
  • Mini Pin Jack (mating pins provided)
  • Output Connector:
  • BNC
  • Input Impedance:
  • 1000 gigohm
  • Output Impedance:
  • 560 ohm
  • Input Range
  • +100mV: DC: +1V
  • Output Range:
  • +10V
  • Gain:
  • X100, X1,000, X10,000
  • Coupling:
  • AC
  • Filter(Hz):
  • Active 20 dB/decade

Low Cut: 1, 10, 300 (other cutoff frequencies available)

High Cut: 300, 3k, 10k (other cutoff frequencies available)

  • Power:
  • Includes PS-1 External Power Supply 115V or 230VAC
  • Size:

Mainframe: 3.5" X 19" Rack Mount,

Headstage: 1.6" high X 5.5" wide X 6" deep

  • Noise:
  • (RMS & Peak to Peak ) see table

Source Resistance = Shorted Input

Bandwidth

RMS

P-P

1Hz -100Hz

0.3uV

1.5uV

1Hz -1KHz

0.7uV

3.5uV

1Hz -10KHz

1.6uV

8.0uV

Source Resistance = 100 Kilohm

Bandwidth

RMS

P-P

1Hz -100Hz

0.55uV

2.75uV

1Hz -1KHz

1.6uV

8.0uV

1Hz -10KHz

3.9uV

19.5uV

Source Resistance = 1 Megohm

Bandwidth

RMS

P-P

1Hz -100Hz

1.6uV

8.0uV

1Hz -1KHz

4.7uV

23.5uV

1Hz -10KHz

11uV

55uV

Source Resistance = 10 Megohm

Bandwidth

RMS

P-P

1Hz -100Hz

5.0uV

25uV

1Hz -1KHz

14uV

70uV

1Hz -10KHz

25uV

125uV

 

Price:

EX-1000

10 Channel Extracellular Amplifier with 10 Channel Headstage and PS-1 External Power Supply

$ 4200.00

Z-I option

Electrode Resistance Test and Current Pump

$ 750.00



See the Holders and Accesories section for extra input pins

 

 Other Extracellular Amplifiers:

  • 2400A Preamplifier Current Pump System
  • EX4-400 4 Channel Differential
  • EX1 1 Channel Differential
  • Amp-8 8 Channel Preamplifier

 

Warning: This equipment is not intended for use in human applications or human experimentation.

[ Home | News | Products | Ordering | Feedback ]

Dagan Corporation Minneapolis MN USA  Email: info@dagan.com 12/10